OxyScout™ Ultra-Trace and OxyScout™ Trace are optical oxygen analyzers for continuous ppm-level O₂ measurement in natural gas. Ultra-Trace covers 0–50 ppm* for low-ppm specifications, and Trace covers 0–1,000 ppm*. Neither model has an electrochemical cell to replace.
A trace oxygen analyzer measures oxygen in parts per million (ppm) by volume. Ambient air is about 20.9% oxygen, or 209,000 ppm, so a small leak of air into a sample line or process stream shows up quickly at ppm levels. OxyScout™ trace models measure with a fully optical sensor and send the reading over 4–20 mA and RS-485 Modbus RTU at the same time.
| Model | Measuring range* | Detection limit* | Accuracy* | Typical application |
|---|---|---|---|---|
| OxyScout Ultra-Trace | 0–50 ppm O₂ | TBD | ±2 ppm or ±5% of reading, whichever is greater | Sales gas, custody transfer and vapor recovery units with a low-ppm specification |
| OxyScout Trace | 0–1,000 ppm O₂ | 50 ppm O₂ | Typically ±2% of reading or ±0.01% O₂ (100 ppm), whichever is greater | Gathering, vapor recovery units and process |
*Based on controlled lab conditions.
Choose a range that puts your oxygen limit well inside the scale, with enough headroom above the limit to see an upset rather than a pegged signal. OxyScout Trace has a 50 ppm detection limit*, so a low-ppm specification calls for OxyScout Ultra-Trace. The pipeline oxygen limits guide matches published limits to each range.
OxyScout Ultra-Trace is the low-ppm model, with a 0–50 ppm O₂ range*.
| Measuring range* | 0–50 ppm O₂ |
|---|---|
| Detection limit* | TBD |
| Accuracy* | ±2 ppm or ±5% of reading, whichever is greater |
| Typical application | Sales gas, custody transfer and vapor recovery units with a low-ppm specification |
*Based on controlled lab conditions.
OxyScout Trace covers 0–1,000 ppm O₂* with a 50 ppm O₂ detection limit*.
| Measuring range* | 0–1,000 ppm O₂ |
|---|---|
| Detection limit* | 50 ppm O₂ |
| Accuracy* | Typically ±2% of reading or ±0.01% O₂ (100 ppm), whichever is greater |
| Typical application | Gathering, vapor recovery units and process |
*Based on controlled lab conditions.
Both trace models share the installation, wiring, firmware and communications of the full OxyScout line. The full specification table also covers options and accessories.
| Measurement | The analyzer uses a fully optical O₂ sensor, so there are no electrochemical cells to replace. |
|---|---|
| Outputs | The outputs are a 4–20 mA loop (current sink) and RS-485 Modbus RTU (2-wire half-duplex). |
| Power | The analyzer runs on 9–36 VDC and draws 35 mA plus the current used by its analog outputs. |
| Process / ambient temperature | Process and ambient temperature is 0 to 50 °C, and storage temperature is −20 to 70 °C. |
| Pressure | Up to 250 PSI input pressure† |
| Humidity / flow | The analyzer is specified for non-condensing conditions, and it has no flow restrictions. |
| Area classification | The analyzer is engineered for Class 1 Division 2 (Zone 2) hazardous areas. |
| Size / weight | The enclosure is 9 in high, 7 in wide and 2.5 in deep, and the analyzer weighs 4 lb. |
| Process connection | The process connection is 1/8 in tube compression. |
| Response time* | Typical response time is under 10 seconds. |
*Based on controlled lab conditions. Standard Sensors Corporation reserves the right to make revisions and/or changes without notice.
†Without Flowmeter installed. Higher ratings if regulator installed upstream.
Most high readings at ppm levels come from air entering the sample system, not from the analyzer. Oxygen moves inward through any imperfect fitting against line pressure, because each gas moves according to its own partial pressure.
The ZeroLeak™ Cell Block on every OxyScout is machined from one solid block, so minimal fittings are required inside the analyzer. Read more about the ZeroLeak™ Cell Block.
A vapor recovery unit works on tank vapor near atmospheric pressure, which is where air gets in. The analyzer mounts on the vapor line between the separator and the VRU compressor.
A gathering system is miles of fittings, packings and welds, and any of them can let air in when the line runs near or below atmospheric. An optical oxygen analyzer on the trunk line reads the field's combined oxygen before the transmission interconnect.
A limit below 1,000 ppm falls in the Trace or Ultra-Trace ranges. For a midstream operator, the analyzer mounts on the pipeline just downstream of the fence line at the receipt point.
Worn rod packing and shaft seals can let air into a compressor that pulls on a low-pressure source. The 0–50 ppm and 0–1000 ppm ranges suit comparison against a tariff limit.
At the wellhead, oxygen is still undiluted and traceable to one well. An optical oxygen analyzer on produced gas at the wellhead or production separator turns a field-wide spec problem into a single work order.
After oxygen removal, the reading shows whether the removal step is working. The range to consider there is Trace, 0–1000 ppm, or Ultra-Trace, 0–50 ppm.
The analyzer uses a fully optical O₂ sensor, so there are no electrochemical cells to replace. OxyScout has no cross-sensitivity or poisoning from H₂S, CO₂ or the other gases that plague galvanic cells. The optical vs electrochemical page compares the two categories.
OxyScout measures oxygen continuously in natural gas, and the reading goes out over 4–20 mA and RS-485 Modbus RTU at the same time.
The analyzer uses a fully optical O₂ sensor, so there are no electrochemical cells to replace.
OxyScout Trace has a detection limit of 50 ppm O₂*, so Ultra-Trace is the model for a low-ppm specification. *Based on controlled lab conditions.
Air trapped in tubing, fittings and regulators bleeds slowly into the flowing sample. On a fresh installation, expect the reading to fall steadily for several hours, and longer on long or complex sample runs, before it reaches a stable low-ppm value.
OxyScout has no cross-sensitivity or poisoning from H₂S, CO₂ or the other gases that plague galvanic cells.
Choose by measuring range. OxyScout Ultra-Trace reads 0–50 ppm. OxyScout Trace reads 0–1,000 ppm with a detection limit of 50 ppm O₂*, so a low-ppm specification calls for Ultra-Trace. *Based on controlled lab conditions.
Choose by measuring range. OxyScout Trace reads 0–1,000 ppm O₂ with a 50 ppm detection limit, and OxyScout Percent reads 0–1, 0–10 or 0–21 %vol O₂ with a 0.005% O₂ (50 ppm) detection limit. *Based on controlled lab conditions.
Zero gas is ultra-high-purity (UHP) nitrogen, 99.999%, certified to less than 1 ppm O₂, through a stainless-steel-diaphragm regulator that has been properly bled.
No. The optical sensor has no cross-sensitivity to H2S. In field testing of the sensing technology, the sensor ran in gas containing high H2S concentration (percent level) with no effect. OxyScout is intended for dry, non-condensing sample gas that contains H2S. For wet sour gas, contact Standard Sensors to review the application suitability. See the oxygen analyzer for H2S-bearing natural gas page.
Book a free 30-minute online meeting to review your application, range and quantities, then we send a quote. You can also call (936) 259-4643 or email [email protected].
Book a 30-minute meeting to review your application and we'll send you a quote after the call.