Intended for dry, non-condensing sample gas that contains H2S.
Last updated:
Produced gas can carry hydrogen sulfide and carbon dioxide. OxyScout™ uses a proprietary, fully optical sensor for continuous oxygen measurement. OxyScout is intended for dry, non-condensing sample gas that contains H2S. Wet sour gas is a different application condition; contact Standard Sensors to review suitability before specifying the analyzer for wet sour service.
The U.S. Energy Information Administration states that wellhead natural gas may contain nonhydrocarbons such as hydrogen sulfide and carbon dioxide. That statement supports the composition sentence above. The cross-sensitivity statements on this page are OxyScout statements and are not taken from that source. Source: EIA, Natural gas explained.
The optical sensor has no cross-sensitivity to H2S. In field testing of the sensing technology, the sensor ran in gas containing high H2S concentration (percent level) with no effect. OxyScout is intended for dry, non-condensing sample gas that contains H2S. For wet sour gas, contact Standard Sensors to review the application suitability.
The percent-level field test describes the sensing technology. That test is not a published H2S percentage or ppm service limit for fittings or other hardware, and this page does not state one.
The sensing technology has been field tested in up to 100% CO2 with no effect on the reading and no damage.
Liquid water offsets the reading temporarily. Once the sample path is purged, the reading recovers. OxyScout is specified for non-condensing sample conditions. If the sample can be wet or condensing in sour service, request an application review.
Organic solvent vapors, chlorine gas, and nitrogen dioxide (NO2) interfere with the sensor. Methane, CO2, and H2S do not interfere.
Ultra-Trace OxyScout-UT (0–50 ppm), Trace OxyScout-T (0–1,000 ppm), and Percent OxyScout-P (0–1 / 0–10 / 0–21 %vol) share the same communications. SKU and MPN are identical for each model. Outputs are a 4–20 mA loop (current sink) and RS-485 Modbus RTU. Power is 9–36 VDC. Input pressure is up to 250 psi without the optional flowmeter. Operating temperature is 0 to 50 °C. The enclosure is IP67. The analyzer weighs 4 lb. OxyScout is engineered for Class 1 Division 2 (Zone 2) hazardous areas. OxyScout is built in the USA, and covered by a 12-month warranty from shipment that includes the sensor.
| Gas or condition | Effect on OxyScout |
|---|---|
| Natural gas, methane (CH4) | No cross-sensitivity |
| Hydrogen sulfide (H2S) | No cross-sensitivity of the optical sensor. In field testing of the sensing technology, high H2S concentration (percent level) had no effect. Intended for dry, non-condensing sample gas that contains H2S. Wet sour gas requires application review. |
| Carbon dioxide (CO2) | No cross-sensitivity. Field tested up to 100%. |
| Liquid water | Temporary offset. Recovers after purging. |
| Other organic solvent vapors | Interferes |
| Chlorine gas, nitrogen dioxide (NO2) | Interferes |
No. The optical sensor has no cross-sensitivity to H2S. In field testing of the sensing technology, the sensor ran in gas containing high H2S concentration (percent level) with no effect. OxyScout is intended for dry, non-condensing sample gas that contains H2S. For wet sour gas, contact Standard Sensors to review the application suitability.
No. The sensing technology has been field tested in up to 100% CO2 with no effect on the reading and no damage.
Organic solvent vapors, chlorine gas, and nitrogen dioxide (NO2) interfere with the sensor. Methane, CO2, and H2S do not interfere.
Liquid water offsets the reading temporarily. Once the sample path is purged, the reading recovers. OxyScout is specified for non-condensing sample conditions.
The sensor is covered by the 12-month warranty. Standard Sensors expects more than one year of service, and longer in clean, dry sample conditions. Long-term life data is still being collected.
A two-point calibration with zero gas and span gas is recommended. A one-point calibration is acceptable. Drift between calibrations depends on the sample rate set and on process quality. Typically, a 1–2 month interval results in low single-digit ppm of oxygen drift across OxyScout models. Drift between calibrations is not the same as detection limit or accuracy.
Air doesn't damage or poison the OxyScout optical sensor. Once process gas is flowing again, the reading typically recovers within minutes. Electrochemical trace O₂ cells can be used up or permanently damaged by extended air exposure.
OxyScout is built in the USA. The standard warranty is 12 months from shipment and includes the sensor. Lead time can be less than one week.
Book a 30-minute meeting to review your application and we'll send you a quote after the call.